Signet 2760 DryLoc pH/ORP Preamplifi er

Wet-Tap Assembly and Installation

The 2760 DryLoc Preamplifi er can be used to connect Signet 3719 Wet-Tap electrodes to a pH or ORP instrument.

1. Select the process fi tting.
• Select a low profi le installation saddle (customer-supplied) with a 1½ in. or 2 in. outlet (depending on 3719 model ordered).
• The 3719 is available with NPT or ISO thread options.
• The 3719 can also be adapted to fi t into tee fi ttings. Refer to the 3719 pH Wet-Tap instructions for detailed information.

2. Install the Wet-Tap and the electrode.
• Install the 3719 into the fi tting, then insert the electrode into the assembly.

3. Attach the preamplifi er to the electrode.
• Attach the 3-2760-11 or -21 Preamplifi er to the DryLoc fi tting at the top of the electrode.

4. Connect the 2760 Preamplifi er to the instrument.
• The 2760 Preamplifi er signal can be routed to any 5700 or 8750 pH/ORP instrument (see Pg. 5).

R LA 1

2

Lubricate O-rings

Lubricate O-rings