beautypg.com

GF Signet 2551 Magmeter Flow Sensor - Blind User Manual

Page 8

background image

8

2551 Magmeter

Pipe Size

(Metric)

Fitting

Type

K-Factor
Gallons

K-Factor
Liters

20 mA=
in GPM

20 mA=
in LPM

POLYPROPYLENE FITTINGS (DIN/ISO, BS, ANSI)

DN15

PPMT005

2192.73

579.32

13.68

51.78

DN20

PPMT007

1327.81

350.81

22.59

85.52

DN25

PPMT010

737.16

194.76

40.70

154.04

DN32

PPMT012

453.46

119.81

66.16

250.41

DN40

PPMT015

275.03

72.66

109.08

412.86

DN50

PPMT020

164.17

43.35

182.74

691.66

PVDF FITTINGS (DIN/ISO, BS, ANSI)

DN15

SFMT005

1946.49

514.26

15.41

58.34

DN20

SFMT007

1158.05

305.96

25.91

98.05

DN25

SFMT010

749.09

197.91

40.05

151.58

DN32

SFMT012

439.51

116.12

68.26

258.36

DN40

SFMT015

248.93

65.77

120.52

456.16

DN50

SFMT020

146.85

38.80

204.30

773.26

PVC FITTINGS (DIN/ISO, BS, ANSI)

DN15

PVMT005

2067.76

546.30

14.51

54.91

DN20

PVMT007

1136.61

300.29

26.39

99.90

DN25

PVMT010

716.52

189.31

41.87

158.47

DN32

PVMT012

446.07

117.85

67.25

254.56

DN40

PVMT015

278.83

73.67

107.59

407.23

DN50

PVMT020

159.36

42.10

188.26

712.55

Plastic Installation Fittings for Metric Pipes:

Polypropylene True Union Tees and Wafers
PVDF True Union Tees
PVC True Union Tees

10. Calibration Data: K-factors and Full Scale Current Values