Burkert Type 8035 User Manual
Page 12
12
Technical data
Type 8025/8035 /
Pulse output
• type
• function
• frequency (f)
• electrical data
• duty cycle
• protection
polarized, potential-free
• NPN/PNP (wiring dependant)
• pulse output, adjustable pulse
value
• 2,5-400 Hz
• 5-36 V DC, 100 mA max.,
voltage drop 2,5 V DC at
100 mA
• 0,5
• galvanically isolated, and
protected against over-
voltages, polarity reversals
and short-circuits
Relay outputs
• operating
• electrical data of the load (non
UL recognized devices)
• electrical data of the load (UL
recognized devices)
• hysteresis, adjustable
thresholds, normally open
• 230 V AC / 3 A
• 30 V AC max. and 42 V peak
or 60 V DC max.
Current output
• response time (10% to 90%)
• max. loop impedance,
12-36 V DC version
• max. loop impedance,
115/230 V AC version
• wiring, version without relays
• wiring, version with relays
4-20 mA, sinking or sourcing
mode (through wiring)
• 5,75 s (default)
• 900 W at 30 V DC, 600 W at
24 V DC, 50 W at 12 V DC
• 800 W
• 2-wire
• 3-wire
Protection against polarity
reversals
yes
6.6.
specifications of the connected
flow sensor
Sensor input
• signal frequency
• pulse signal (Hall)
• sinus signal (coil)
• 2,5 to 400 Hz
• NPN, open collector
• typical sensitivity of 35 mV peak-
peak, at 252 Hz
Sensor output
• power supply
• 10-34 V DC (V+ subtract 2
V DC), 1 mA max.
English