beautypg.com

Power meter (frequency-selective) – Atec R-S-CMU200 User Manual

Page 9

background image

Version 09.00, October 2008

R&S

®

CMU200 Universal Radio Communication Tester

9

Power meter (frequency-selective)

Frequency range

10 MHz to 2700 MHz

Frequency resolution

0.1

Hz

Resolution bandwidths

10 Hz to 1 MHz in 1/2/3/5 steps

Level range

continuous power

3

10 MHz to 2200 MHz

–40 dBm to +47 dBm (50 W)

2200 MHz to 2700 MHz

–34 dBm to +47 dBm (50 W)

RF 1

peak envelope power

4

(PEP)

+53 dBm (200 W)

continuous power

10 MHz to 2200 MHz

–54 dBm to +33 dBm (2 W)

2200 MHz to 2700 MHz

–48 dBm to +33 dBm (2 W)

RF 2

peak envelope power

4

(PEP)

+39 dBm (8 W)

continuous power and PEP

10 MHz to 2200 MHz

–80 dBm to 0 dBm

RF 4 IN

2200 MHz to 2700 MHz

–74 dBm to 0 dBm

Level uncertainty

in temperature range +20 °C to +35 °C

50 MHz to 2200 MHz

<0.5 dB

RF 1, RF 2

2200 MHz to 2700 MHz

<0.7 dB

50 MHz to 2200 MHz

<0.7 dB

RF 4 IN

2200 MHz to 2700 MHz

<0.9 dB

Level uncertainty

in temperature range +5 °C to +45 °C

50 MHz to 2200 MHz

<1.0 dB

RF 1, RF 2

2200 MHz to 2700 MHz

<1.0 dB

50 MHz to 2200 MHz

<1.0 dB

RF 4 IN

2200 MHz to 2700 MHz

<1.1 dB

in manual mode

0.1 dB

Level resolution

in remote control mode

0.01 dB

typical values after 1 h warm-up time

input level

≥ –40 dBm

<0.01 dB

RF level measurement repeatability

input level < –40 dBm

<0.03 dB