Cmu-b96 additional rf generator option, Path 2 for wcdma – Atec R-S-CMU200 User Manual
Page 33

Version 09.00, October 2008
R&S
®
CMU200 Universal Radio Communication Tester
33
R&S
®
CMU-B96 additional RF generator option
Path 2 for WCDMA
Standard
3GPP
FDD
non-signaling mode
Frequency range
728 MHz to 768 MHz
869 MHz to 894 MHz
925 MHz to 960 MHz
1475.9 MHz to 1500.9 MHz
1805 MHz to 1880 MHz
1930 MHz to 1990 MHz
2110 MHz to 2170 MHz
2180 MHz to 2200 MHz
WCDMA signaling mode
band 1
2110 MHz to 2170 MHz
band 2
1930 MHz to 1990 MHz
band 3
1805 MHz to 1880 MHz
band 4
2110 MHz to 2155 MHz
band 5
869 MHz to 894 MHz
band 6
875 MHz to 885 MHz
band 8
925 MHz to 960 MHz
band 9
1844.9 MHz to 1879.9 MHz
band 10
2110 MHz to 2170 MHz
band 11
1475.9 MHz to 1500.9 MHz
band 12
728 MHz to 746 MHz
band 13
746 MHz to 756 MHz
band 14
758 MHz to 768 MHz
Frequency range
band S
2180 MHz to 2200 MHz
Frequency resolution
2.5
kHz
Frequency uncertainty
same as timebase,
see base unit specifications
Error vector magnitude (EVM)
global EVM for DL RMC in line with 3GPP
TS 34.121 C3.1 to C3.4 with
DPCH/CPICH = 0 dB
<8 %, rms
Output level range
without R&S
®
CMU-U99
–115 dBm to –82 dBm
RF 1
with R&S
®
CMU-U99
–103 dBm to –70 dBm
RF 2
–103 dBm to –70 dBm
Output level resolution
RF 1 and RF 2
0.1 dB
Output level range
overrange mode
without R&S
®
CMU-U99
–110 dBm to –38 dBm
RF 1
with R&S
®
CMU-U99
–90 dBm to –24 dBm
RF 2
–90 dBm to –24 dBm
RF 3 OUT
–70 dBm to –1 dBm
Output level resolution
for overrange mode
1 dB