Modulation analyzer, Timing measurement – Atec R-S-CMU200 User Manual
Page 40
40 R&S
®
CMU200 Universal Radio Communication Tester
Version 09.00, October 2008
Power meter (frequency-selective) and power versus time
filter definition: passband
Filter Bandwidth
→ wide
2.0 MHz
Measurement bandwidth
Filter Bandwidth
→ narrow 1.3
MHz
Reference level for full dynamic range
GFSK signal
RF 1
0 dBm to +41 dBm
RF 2
–14 dBm to +33 dBm
RF 4 IN
–32 dBm to 0 dBm
Dynamic range
Filter Bandwidth
→ wide
>55 dB, rms
Level uncertainty
in temperature range +20 °C to +35 °C
RF 1, RF 2
from full scale setting down to –25 dB
<0.7 dB
RF 4 IN
from full scale setting down to –25 dB
<0.9 dB
Level uncertainty
in temperature range +5 °C to +45 °C
RF1, RF2
from full scale setting down to –25 dB
<1.0 dB
RF4IN
from full scale setting down to –25 dB
<1.1 dB
in manual mode
0.1 dB
Level resolution
in remote control mode
0.01 dB
Modulation analyzer
filter definition: passband
Filter Bandwidth
→ wide
2.0 MHz
Measurement bandwidth
Filter Bandwidth
→ narrow 1.3
MHz
Level range
GFSK signal
RF 1, RF 2, RF 4 IN
from full scale setting down to –25 dB
Total measurement range for frequency
offset and frequency deviation
–250 kHz to +250 kHz
Frequency offset uncertainty in
preamble
for deviation
≤ 160 kHz
≤2 kHz
for 100 kHz < deviation
≤ 200 kHz
11110000 pattern
≤2 %
Frequency deviation uncertainty in
payload
10101010 pattern
≤4 %
measured in burst related to frequency
offset value in preamble
10101010 pattern
max.
≤2 kHz
Frequency drift uncertainty
typ.
≤1 kHz
in manual mode
1 kHz
Frequency resolution
in remote control mode
1 Hz
Timing measurement
Range
±20
μs
Resolution
≤0.25 μs
Uncertainty
≤0.25 μs + resolution