beautypg.com

Rf analyzer, Influence on rf interface, Power meter (frequency-selective) – Atec R-S-CMU200 User Manual

Page 15

background image

Version 09.00, October 2008

R&S

®

CMU200 Universal Radio Communication Tester

15

Output level range

8PSK

RF 1

without R&S

®

CMU-U99

with R&S

®

CMU-U99

–115 dBm to –76 dBm
–103 dBm to –64 dBm

RF 2

–103 dBm to –64 dBm

Output level range

GMSK overrange mode

RF 1

without R&S

®

CMU-U99

with R&S

®

CMU-U99

–110 dBm to –28 dBm
–90 dBm to –14 dBm

RF 2

–90 dBm to –14 dBm

RF 3 OUT

–70 dBm to +9 dBm

Output level range

8PSK overrange mode

RF 1

without R&S

®

CMU-U99

with R&S

®

CMU-U99

–110 dBm to –32 dBm
–90 dBm to –18 dBm

RF 2

–90 dBm to –18 dBm

RF 3 OUT

–70 dBm to +5 dBm

Output level resolution

1

dB

Influence on RF interface

Reduced input level range

if R&S

®

CMU-B96 is installed

RF 1

continuous input power

max. 2 W

RF level uncertainty

Use of the R&S

®

CMU-B96 in overrange

mode may influence all RF signal levels
and their quality.

RF analyzer

GSM400 band

450 MHz to 458 MHz
478 MHz to 486 MHz

GSM850 band

824 MHz to 849 MHz

GSM900 band

876 MHz to 915 MHz

GSM1800 band

1710 MHz to 1785 MHz

Frequency range

GSM1900 band

1850 MHz to 1910 MHz

Power meter (frequency-selective)

Level range

continuous power

3

–40 dBm to +47 dBm (50 W)

RF 1

peak envelope power

4

(PEP)

+53 dBm (200 W)

continuous power

–54 dBm to +33 dBm (2 W)

RF 2

peak envelope power

4

(PEP)

+39 dBm (8 W)

RF 4 IN

continuous power and PEP

–80 dBm to 0 dBm

in temperature range +20 °C to +35 °C

<0.5 dB

Level uncertainty

in temperature range +5 °C to +45 °C

<0.7 dB

in manual mode

0.1 dB

Level resolution

in remote control mode

0.01 dB

Measurement bandwidth

selectable

500 kHz or 600 kHz