Atec R-S-CMU200 User Manual
Page 32
32 R&S
®
CMU200 Universal Radio Communication Tester
Version 09.00, October 2008
manual mode
0.1 dB
Output level resolution
remote mode
0.01 dB
Output level uncertainty
in temperature range +20 °C to +35 °C,
no overranging
output level
≥ –120 dBm
f < 2200 MHz
<0.6 dB
RF 1, RF 2
f
≥ 2200 MHz
<0.8 dB
output level
≥ –80 dBm
f < 2200 MHz
<0.8 dB
RF 3 OUT
f
≥ 2200 MHz
<1.0 dB
Output level uncertainty
in temperature range +5 °C to +45 °C,
no overranging
output level
≥ –120 dBm
f < 2200 MHz
<0.9 dB
RF 1, RF 2
f
≥ 2200 MHz
<1.5 dB
output level
≥ –80 dBm
f < 2200 MHz
<1.0 dB
RF 3 OUT
f
≥ 2200 MHz
<1.5 dB
Output level setting
setting reference
relative to CPICH or
total output power
non-signaling mode
P-CPICH, P-SCH, S-SCH, P-CCPCH,
PICH, DPCH, OCNS, HS-SCCH,
HS-PDSCH, E-AGCH
–35 dB to +15 dB relative to CPICH
Channel levels
E-RGCH/E-HICH
–35 dB to +12 dB relative to CPICH
signaling mode
P-CPICH, P-SCH, S-SCH, P-CCPCH,
S-CCPCH, PICH, AICH, DPCH, OCNS,
HS-SCCH, HS-PDSCH, E-AGCH
–35 dB to +15 dB relative to CPICH
Channel levels
E-RGCH/E-HICH
–35 dB to +12 dB relative to CPICH
Signal quality
Error vector magnitude (EVM)
global EVM for DL RMC in line with 3GPP
TS 34.121 C3.1 to C3.4 with
DPCH/CPICH = 0 dB
<8 %, rms
Signal quality
16QAM
Error vector magnitude (EVM)
global EVM for 16QAM
reference setup: 3GPP TS34.121 FRC
H-Set3 for 16QAM
<8 %, rms