R&S

®

CMU200 Universal Radio Communication Tester

19

Power meter (frequency-selective)

Level range

see base unit specifications

Level uncertainty

see base unit specifications

Power versus time measurement

Reference level for full dynamic range

RF 1

+4 dBm to +47 dBm

RF 2

–10 dBm to +33 dBm

RF 4 IN

–28 dBm to –6 dBm

Dynamic range

Filter

100 kHz, rms

RF Attenuation

Low Noise

>74 dB

result > –40 dB

<0.1 dB

Relative measurement uncertainty

–60 dB

≤ result ≤ –40 dB

<0.5 dB

Residual leakage power level

<–65

dBm

Adjacent channel power measurement

first adjacent channel

>45 dB

Dynamic range

second and third adjacent channel

>55 dB