beautypg.com

Atec Agilent-16700 Series User Manual

Page 99

background image

99

[1] Minimum eye width in system under test
[2] The choice of probe can limit system performance. Select a probe rated at 600 Mb/s or greater to maintain system bandwidth.
[3] Sample positions are independently adjustable for each data channel input. A negative sample position causes the input to be synchronously sampled by that amount before

each active clock edge. A positive sample position causes the input to be synchronously sampled by that amount after each active clock edge. A sampling position of zero
causes the input to be synchronously sampled coincident with each clock edge.

[4] Use of eye finder is recommended in 600 Mb/s mode
[5] With time or state tags on and all pods assigned, memory depth is half the maximum memory depth. With time or state tags on and one pod (34 channels) unassigned, the

memory depth is full.

[6] Independent analyzers may be either state or timing. When the 600 Mb/s state mode is selected, only one analyzer may be used.
[7] In the 300 Mb/s state mode, the total number of clocks and qualifiers is 4. All clock and qualifier inputs must be on the master module.
[8] Tested with input signal Vh = 0.9 V, Vl = -1.7 V, slew rate = 1 V/ns, threshold = -1.3 V
[9] Calculated from rise time
Items marked with an asterisk * are specifications. All others are characteristics.
"Typical" represents the average or median value of the parameter based on measurements from a significant number of units.

State/Timing Modules Specifications and
Characteristics

Agilent 16753A, 16754A, 16755A, 16756A
Supplemental Specifications* and Characteristics (continued)

State (synchronous) analysis mode (continued)

300 Mb/s State Mode

600 Mb/s State Mode

Maximum trigger sequence speed

300 MHz

600 MHz

Maximum trigger sequence levels

16

16

Trigger sequence level branching

Arbitrary 4-way "IF/THEN/ELSE"

2-way "IF/THEN/ELSE"

Trigger position

Start, center, end, or user -defined

Start, center, end, or user -defined

Trigger resources

16 patterns evaluated as =, =/, >, >=, <, <=

14 patterns evaluated as =, =/, >, >=, <, <=

14 double-bounded ranges evaluated as in range,

7 double-bounded ranges evaluated as in range,

not in range

not in range

2 timers per module

1 occurrence counter per sequence level

2 global counters

4 flags

1 occurrence counter per sequence level
4 flags

Trigger resource conditions

Arbitrary Boolean combinations

Arbitrary Boolean combinations

Trigger actions

Goto

Goto

Trigger and fill memory

Trigger and fill memory

Trigger and Goto
Store/don't store sample
Turn on/off default storing
Timer start/stop/pause/resume
Global counter increment/decrement/reset
Occurrence counter reset
Flag set/clear

Store qualification

Default (global) and per sequence level

Default (global)

Maximum global counter

2E+24

N/A

Maximum occurrence counter

2E+24

2E+24

Maximum pattern/range width

32 bits

32 bits

Timers range

40 ns to 2199 seconds

N/A

Timer resolution

2 ns

N/A

Timer accuracy

+/- (5 ns +0.01%)

N/A

Timer reset latency

40 ns

N/A