Atec Agilent-16700 Series User Manual
Page 97
97
Timing Zoom
Timing analysis sample rate
4 GHz
Time interval accuracy, within a pod pair
+/- (750 ps + 0.01% of time interval reading)
Time interval accuracy, between pod pairs
+/- (1.5 ns + 0.01% of time interval reading)
Memory depth
64 K samples
Trigger position
Start, center, end, or user defined
Minimum data pulse width
750 ps
State (synchronous) analysis mode
300 Mb/s State Mode
600 Mb/s State Mode
tWidth*[1,2] with E5378A, E5379A, E5387A,
1 ns*, 600 ps typical
1 ns*, 600 ps typical
or E5390A probes
tWidth[1] with E5380A or E5382A probes
1.5 ns, 1 ns typical
1.5 ns, 1 ns typical
tSetup
0.5 tWidth
0.5 tWidth
tHold
0.5 tWidth
0.5 tWidth
tSample range [3]
-4 ns to +4 ns
-4 ns to +4 ns
tSample adjustment resolution
80 ps (typical)
80 ps (typical)
tSample accuracy, manual adjustment
+/- 300 ps
+/- 300 ps [4]
State/Timing Modules Specifications and
Characteristics
Individual
Data Channel
vHeight
tWidth
tSetup
Sampling
Position
Clock Channel
Note (1)
tHold
vThreshold*
*User Adjustable
—0V—
tSample*
Data Eye
[1] Minimum eye width in system under test
[2] The choice of probe can limit system performance. Select a probe rated at 600 Mb/s or greater to maintain system bandwidth.
[3] Sample positions are independently adjustable for each data channel input. A negative sample position causes the input to be synchronously sampled by that amount before
each active clock edge. A positive sample position causes the input to be synchronously sampled by that amount after each active clock edge. A sampling position of zero
causes the input to be synchronously sampled coincident with each clock edge.
[4] Use of eye finder is recommended in 600 Mb/s mode
Items marked with an asterisk * are specifications. All others are characteristics.
"Typical" represents the average or median value of the parameter based on measurements from a significant number of units.
Agilent 16753A, 16754A, 16755A, 16756A
Supplemental Specifications* and Characteristics
Figure 6.14. Data Sampling.