beautypg.com

Atec Agilent-16700 Series User Manual

Page 101

background image

101

[1] Minimum eye width in system under test
[2] The choice of probe can limit system performance. Select a probe rated at 600 Mb/s or greater to maintain system bandwidth.
[3] Sample positions are independently adjustable for each data channel input. A negative sample position causes the input to be synchronously sampled by that amount before

each active clock edge. A positive sample position causes the input to be synchronously sampled by that amount after each active clock edge. A sampling position of zero
causes the input to be synchronously sampled coincident with each clock edge.

[4] Use of eye finder is recommended in 600 Mb/s mode
[5] With time or state tags on and all pods assigned, memory depth is half the maximum memory depth. With time or state tags on and one pod (34 channels) unassigned, the

memory depth is full.

[6] Independent analyzers may be either state or timing. When the 600 Mb/s state mode is selected, only one analyzer may be used.
[7] In the 300 Mb/s state mode, the total number of clocks and qualifiers is 4. All clock and qualifier inputs must be on the master module.
[8] Tested with input signal Vh = 0.9 V, Vl = -1.7 V, slew rate = 1 V/ns, threshold = -1.3 V
[9] Calculated from rise time
Items marked with an asterisk * are specifications. All others are characteristics.
"Typical" represents the average or median value of the parameter based on measurements from a significant number of units.

State/Timing Modules Specifications and
Characteristics

Agilent 16753A, 16754A, 16755A, 16756A
Supplemental Specifications* and Characteristics (continued)

Timing (asynchronous) analysis mode

Conventional timing

Transitional timing

Trigger resources

16 patterns evaluated as =, =/, >, >=, <, <=

16 patterns evaluated as =, =/, >, >=, <, <=

14 double-bounded ranges evaluated as in range,

14 double-bounded ranges evaluated as in range,

not in range

not in range

3 edge/glitch — per pod pair

3 edge/glitch — per pod pair

2 timers per module

2 timers per module

2 global counters

2 global counters

1 occurrence counter per sequence level

1 occurrence counter per sequence level

4 flags

4 flags

Trigger resource conditions

Arbitrary Boolean combinations

Arbitrary Boolean combinations

Trigger actions

Goto

Goto

Trigger and fill memory

Trigger and fill memory

Trigger and Goto

Trigger and Goto

Turn on/off default storing

Turn on/off default storing

Timer start/stop/pause/resume

Timer start/stop/pause/resume

Global counter increment/decrement/reset

Global counter increment/decrement/reset

Occurrence counter reset

Occurrence counter reset

Flag set/clear

Flag set/clear

Maximum global counter

2E+24

2E+24

Maximum occurrence counter

2E+24

2E+24

Maximum pattern/range width

32 bits

32 bits

Timer value range

40 ns to 2199 seconds

40 ns to 2199 seconds

Timer resolution

2 ns

2 ns

Timer accuracy

+/- (5 ns +0.01%)

+/- (5 ns +0.01%)

Greater than duration

3.33 ns to 55 ms in 3.3 ns increments

3.33 ns to 55 ms in 3.3 ns increments

Less that duration

6.67 ns to 55 ms in 3.3 ns increments

6.67 ns to 55 ms in 3.3 ns increments

Timer reset latency

40 ns

40 ns