Atec Agilent-16700 Series User Manual
Page 101
101
[1] Minimum eye width in system under test
[2] The choice of probe can limit system performance. Select a probe rated at 600 Mb/s or greater to maintain system bandwidth.
[3] Sample positions are independently adjustable for each data channel input. A negative sample position causes the input to be synchronously sampled by that amount before
each active clock edge. A positive sample position causes the input to be synchronously sampled by that amount after each active clock edge. A sampling position of zero
causes the input to be synchronously sampled coincident with each clock edge.
[4] Use of eye finder is recommended in 600 Mb/s mode
[5] With time or state tags on and all pods assigned, memory depth is half the maximum memory depth. With time or state tags on and one pod (34 channels) unassigned, the
memory depth is full.
[6] Independent analyzers may be either state or timing. When the 600 Mb/s state mode is selected, only one analyzer may be used.
[7] In the 300 Mb/s state mode, the total number of clocks and qualifiers is 4. All clock and qualifier inputs must be on the master module.
[8] Tested with input signal Vh = 0.9 V, Vl = -1.7 V, slew rate = 1 V/ns, threshold = -1.3 V
[9] Calculated from rise time
Items marked with an asterisk * are specifications. All others are characteristics.
"Typical" represents the average or median value of the parameter based on measurements from a significant number of units.
State/Timing Modules Specifications and
Characteristics
Agilent 16753A, 16754A, 16755A, 16756A
Supplemental Specifications* and Characteristics (continued)
Timing (asynchronous) analysis mode
Conventional timing
Transitional timing
Trigger resources
16 patterns evaluated as =, =/, >, >=, <, <=
16 patterns evaluated as =, =/, >, >=, <, <=
14 double-bounded ranges evaluated as in range,
14 double-bounded ranges evaluated as in range,
not in range
not in range
3 edge/glitch — per pod pair
3 edge/glitch — per pod pair
2 timers per module
2 timers per module
2 global counters
2 global counters
1 occurrence counter per sequence level
1 occurrence counter per sequence level
4 flags
4 flags
Trigger resource conditions
Arbitrary Boolean combinations
Arbitrary Boolean combinations
Trigger actions
Goto
Goto
Trigger and fill memory
Trigger and fill memory
Trigger and Goto
Trigger and Goto
Turn on/off default storing
Turn on/off default storing
Timer start/stop/pause/resume
Timer start/stop/pause/resume
Global counter increment/decrement/reset
Global counter increment/decrement/reset
Occurrence counter reset
Occurrence counter reset
Flag set/clear
Flag set/clear
Maximum global counter
2E+24
2E+24
Maximum occurrence counter
2E+24
2E+24
Maximum pattern/range width
32 bits
32 bits
Timer value range
40 ns to 2199 seconds
40 ns to 2199 seconds
Timer resolution
2 ns
2 ns
Timer accuracy
+/- (5 ns +0.01%)
+/- (5 ns +0.01%)
Greater than duration
3.33 ns to 55 ms in 3.3 ns increments
3.33 ns to 55 ms in 3.3 ns increments
Less that duration
6.67 ns to 55 ms in 3.3 ns increments
6.67 ns to 55 ms in 3.3 ns increments
Timer reset latency
40 ns
40 ns