96
State/Timing Modules Specifications and
Characteristics
Probes for 16753A, 16754A, 16755A, 16756A, 16760A
Supplemental Specifications* and Characteristics (continued)
Probes
E5387A Differential Soft Touch E5390A Single-Ended Soft Touch
Input resistance and
Refer to figure 6.13
Refer to figure 6.13
capacitance
Maximum state data
1.5 Gb/s
1.5 Gb/s
rate supported
Mating connector
None required [5]
None required [5]
Minimum voltage swing
Vin
+ - V
in
- >= 200 mV p-p
250 mV p-p
Input dynamic range
-3 Vdc to +5 Vdc
-3 Vdc to +5 Vdc
Threshold accuracy
+/- (30 mV + 2% of setting)*
+/- (30 mV + 2% of setting) [3]
Threshold range
-3.0 V to +5.0 V
-3.0 V to +5.0 V
User-supplied threshold
N/A
-3.0 V to +5.0 V
input range
User-supplied threshold
N/A
>= 100K ohms
input resistance
Threshold control options
If operated single-ended
• User-provided input
(minus inputs grounded),
• Adjustable from user
the threshold can be adjusted
interface
from the user interface
Maximum nondestructive +/-40 Vdc
+/-40 Vdc
input voltage
Maximum input slew rate
5 V/ns
5 V/ns
Clock input
Differential
Differential
Number of inputs [4]
17 (16 data and 1 clock/data) 34 (32 data and 2 clock/data)
* All specifications noted by an asterisk are the performance standards against which the product is tested.
[1] A support shroud, Agilent part number 16760-02302 (for boards up to 0.062" thick) or 16760-02303 (for boards up to 0.120" thick) is recommended.
A kit of 5 shrouds and 5 connectors is available as Agilent part number 16760-68702 (for boards up to 0.062" thick) or 16760-68703 (for boards up to 0.120" thick).
[2] A kit of 5 Amp Mictor connectors and 5 support shrouds is available, Agilent part number E5346-68701. A support shroud is available separately, Agilent part number E5346-44701.
[3] If operated single-ended (minus inputs grounded), the threshold can be adjusted from the user interface.
[4] Refer to specifications on specific modes of operation for details on how inputs can be used.
[5] Soft touch probes use a retention module attached to the target PC board. A kit of 5 retention modules is included with each probe. Additional kits of 5 retention modules
can be ordered using Agilent part number E5387-68701.
Figure 6.13. E5387A and E5390A Probe Load Model.