beautypg.com

Atec Agilent-16700 Series User Manual

Page 96

background image

96

State/Timing Modules Specifications and
Characteristics

Probes for 16753A, 16754A, 16755A, 16756A, 16760A
Supplemental Specifications* and Characteristics (continued)

Probes

E5387A Differential Soft Touch E5390A Single-Ended Soft Touch

Input resistance and

Refer to figure 6.13

Refer to figure 6.13

capacitance

Maximum state data

1.5 Gb/s

1.5 Gb/s

rate supported

Mating connector

None required [5]

None required [5]

Minimum voltage swing

Vin

+ - V

in

- >= 200 mV p-p

250 mV p-p

Input dynamic range

-3 Vdc to +5 Vdc

-3 Vdc to +5 Vdc

Threshold accuracy

+/- (30 mV + 2% of setting)*

+/- (30 mV + 2% of setting) [3]

Threshold range

-3.0 V to +5.0 V

-3.0 V to +5.0 V

User-supplied threshold

N/A

-3.0 V to +5.0 V

input range

User-supplied threshold

N/A

>= 100K ohms

input resistance

Threshold control options

If operated single-ended

• User-provided input

(minus inputs grounded),

• Adjustable from user

the threshold can be adjusted

interface

from the user interface

Maximum nondestructive +/-40 Vdc

+/-40 Vdc

input voltage

Maximum input slew rate

5 V/ns

5 V/ns

Clock input

Differential

Differential

Number of inputs [4]

17 (16 data and 1 clock/data) 34 (32 data and 2 clock/data)

* All specifications noted by an asterisk are the performance standards against which the product is tested.
[1] A support shroud, Agilent part number 16760-02302 (for boards up to 0.062" thick) or 16760-02303 (for boards up to 0.120" thick) is recommended.

A kit of 5 shrouds and 5 connectors is available as Agilent part number 16760-68702 (for boards up to 0.062" thick) or 16760-68703 (for boards up to 0.120" thick).

[2] A kit of 5 Amp Mictor connectors and 5 support shrouds is available, Agilent part number E5346-68701. A support shroud is available separately, Agilent part number E5346-44701.
[3] If operated single-ended (minus inputs grounded), the threshold can be adjusted from the user interface.
[4] Refer to specifications on specific modes of operation for details on how inputs can be used.
[5] Soft touch probes use a retention module attached to the target PC board. A kit of 5 retention modules is included with each probe. Additional kits of 5 retention modules

can be ordered using Agilent part number E5387-68701.

Figure 6.13. E5387A and E5390A Probe Load Model.