Atec Agilent-16700 Series User Manual

Page 102

background image

102

[1] Minimum eye width in system under test
[2] The choice of probe can limit system performance. Select a probe rated at 600 Mb/s or greater to maintain system bandwidth.
[3] Sample positions are independently adjustable for each data channel input. A negative sample position causes the input to be synchronously sampled by that amount before

each active clock edge. A positive sample position causes the input to be synchronously sampled by that amount after each active clock edge. A sampling position of zero
causes the input to be synchronously sampled coincident with each clock edge.

[4] Use of eye finder is recommended in 600 Mb/s mode
[5] With time or state tags on and all pods assigned, memory depth is half the maximum memory depth. With time or state tags on and one pod (34 channels) unassigned, the

memory depth is full.

[6] Independent analyzers may be either state or timing. When the 600 Mb/s state mode is selected, only one analyzer may be used.
[7] In the 300 Mb/s state mode, the total number of clocks and qualifiers is 4. All clock and qualifier inputs must be on the master module.
[8] Tested with input signal Vh = 0.9 V, Vl = -1.7 V, slew rate = 1 V/ns, threshold = -1.3 V
[9] Calculated from rise time
Items marked with an asterisk * are specifications. All others are characteristics.
"Typical" represents the average or median value of the parameter based on measurements from a significant number of units.

State/Timing Modules Specifications and
Characteristics

Agilent 16753A, 16754A, 16755A, 16756A
Supplemental Specifications* and Characteristics (continued)

Eye scan mode

Equivalent rise time

150 ps

Equivalent bandwidth [9]

2.33 GHz

Sample position range relative to clock

-5 ns to +5 ns

Sample (time) position resolution

10 ps

Sample (time) position accuracy

+/- (50 ps + 0.01 * sample position)

Number of channels

68 * (number of modules) - 1

Input dynamic range

-3.0 Vdc to +5.0 Vdc

Threshold range

-3.0 Vdc to +5.0 Vdc

Threshold resolution

1 mV

Threshold accuracy

+/- (30 mV + 2% of setting)

Minimum detectable pulse width at
minimum signal amplitude

600 ps

Jitter

40 ps RMS

Noise floor

40 mV p-p

Channel-to-channel skew,
maximum between any two channels

100 ps