[1] Minimum eye width in system under test
[2] The choice of probe can limit system performance. Select a probe rated at 600 Mb/s or greater to maintain system bandwidth.
[3] Sample positions are independently adjustable for each data channel input. A negative sample position causes the input to be synchronously sampled by that amount before

each active clock edge. A positive sample position causes the input to be synchronously sampled by that amount after each active clock edge. A sampling position of zero
causes the input to be synchronously sampled coincident with each clock edge.

[4] Use of eye finder is recommended in 600 Mb/s mode
[5] With time or state tags on and all pods assigned, memory depth is half the maximum memory depth. With time or state tags on and one pod (34 channels) unassigned, the

memory depth is full.

[6] Independent analyzers may be either state or timing. When the 600 Mb/s state mode is selected, only one analyzer may be used.
[7] In the 300 Mb/s state mode, the total number of clocks and qualifiers is 4. All clock and qualifier inputs must be on the master module.
[8] Tested with input signal Vh = 0.9 V, Vl = -1.7 V, slew rate = 1 V/ns, threshold = -1.3 V
[9] Calculated from rise time
Items marked with an asterisk * are specifications. All others are characteristics.
"Typical" represents the average or median value of the parameter based on measurements from a significant number of units.

State/Timing Modules Specifications and
Characteristics

Agilent 16753A, 16754A, 16755A, 16756A
Supplemental Specifications* and Characteristics (continued)

Timing (asynchronous) analysis mode

Conventional timing

Transitional timing

Sample rate on all channels

600 MHz

600 MHz

Sample rate in half channel mode

1200 MHz

N/A

Number of channels

68 x (number of modules)

For sample rates <600 MHz:
68 x (number of modules)
For 600 MHz sample rate:
68 x (number of modules) - 34

Maximum channels on a single time base and trigger 340

340

Number of independent analyzers [6]

2

2

Sample period (half channel)

833 ps

N/A

Sample period (full channel)

1.67 ns

1.67 ns

Minimum data pulse width

1 sample period + 500 ps

1 sample period + 500 ps

Time interval accuracy

+/- (1 sample period + 1.25 ns +

+/- (1 sample period + 1.25 ns +

0.01% of time interval reading)

0.01% of time interval reading)

Memory depth in full channel mode

16753A: 1 M

16753A: 1 M

16754A: 4 M

16754A: 4 M

16755A: 16 M

16755A: 16 M

16756A: 64 M

16756A: 64 M

Memory depth in half channel mode

16753A: 2 M

N/A

16754A: 8 M
16755A: 32 M
16756A: 128 M

Maximum trigger sequence speed

300 MHz

300 MHz

Maximum trigger sequence levels

16

16

Trigger sequence level branching

Arbitrary 4-way "IF/THEN/ELSE"

Arbitrary 4-way "IF/THEN/ELSE"

Trigger position

Start, center, end, or user -defined

Start, center, end, or user -defined