2 measured value and status display, Measured value and status display -2 – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual
Page 74
Operation
System 182
4 - 2
ba76022e01
12/2011
4.2
Measured value and status display
With the
m
key you switch to the last selected measured value and
status display from any operating situation. Entries that are not
completed are ignored while doing so.
By pressing
m
once again you cyclically switch between further
display options.
Example:
Display options
with two connected
sensors
Operating notes
Select a sensor in the double display with
w
. The selected sensor
is displayed in reverse video. The number and name of the selected
sensor are displayed in the header. In the single display, the sensor
being displayed is always the selected sensor at the same time.
Start a calibration procedure for the selected sensor with
c
.
Starting a calibration procedure from the display of interfaces is not
possible.
Big double display:
Main measured parameter only.
Detailed double display:
Main and secondary measured
parameter.
Single display.
The display switches between the
connected sensors every
3 seconds.
Display of the outputs.
The current states of the relays
and current outputs and the date
and time are displayed.