8 connecting the voltage supply, 1 iq sensor net (line power version), Connecting the voltage supply -30 3.8.1 – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual
Page 50
Installation
System 182
3 - 30
ba76022e01
12/2011
3.8
Connecting the voltage supply
Note
The two following paragraphs describe how to connect both models of
the IQ S
ENSOR
N
ET
Universal Transmitter to the voltage supply. How
to connect additional power supply modules is described in the
operating manual of the respective power supply module.
3.8.1
IQ S
ENSOR
N
ET
(Line power version)
Warning
If the power supply is incorrectly connected, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
The IQ S
ENSOR
N
ET
Universal Transmitter may only be
connected by a trained electrician.
The connection of the IQ S
ENSOR
N
ET
Universal Transmitter to
the power supply may only be carried out when it is not
carrying any voltage.
The power supply must fulfill the specifications given on the
nameplate and in chapter 9 T
.
When installed in a building, a switch or power switch must be
provided as an interrupt facility for the System 182.
The interrupt facility must
– be installed in the vicinity of the IQ S
ENSOR
N
ET
Universal
Transmitter, easily accessible by the user, and
– be labeled as the interrupt facility for the
IQ S
ENSOR
N
ET
Universal Transmitter.
After the IQ S
ENSOR
N
ET
Universal Transmitter has been
installed, it may only be opened if the line voltage has been
switched off beforehand.
Materials required
Wire end sleeves, suitable for the power line, with suitable crimping
tool
1 x screwed cable gland with sealing ring (scope of delivery of the
Universal Transmitter).
Tools
Cable stripping knife
Wire stripper
Phillips screwdriver
Small screwdriver.