8 indexes, 1 explanation of the messages, 1 error messages – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual
Page 221: 2 info messages, Indexes -1, Explanation of the messages -1 8.1.1, Error messages -1, Info messages -1, 8indexes
MIQ/CR3; DIQ/CR3
Indexes
8 - 1
ba76032e01
01/2012
8
Indexes
8.1
Explanation of the messages
In this chapter you will find a list with all the message codes and
corresponding message texts that may occur in the log book of the
IQ S
ENSOR
N
ET
system for the MIQ/CR3 output module.
Note
Information about
Contents and structure of the log book and
Structure of the message code
can be found in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
Note
All message codes of the output module MIQ/CR3 or DIQ/CR3 end
with the number "411".
8.1.1
Error messages
8.1.2
Info messages
The MIQ/CR3 or DIQ/CR3 output module does not send any info
messages.
Message code
Message text
EA4411
Current output range undercut
* Check process
* Check settings and, if necessary, change them
EA5411
Current output range exceeded
* Check process
* Check settings and, if necessary, change them
EI3411
Burden resistor too large (> 500 Ohm) or current loop interrupted
* Check burden, terminal connections and connection lines