2 system monitoring, See section 4.5.2 – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual
Page 189
MIQ/CR3; DIQ/CR3
Settings
4 - 17
ba76032e01
01/2012
4.5.2
System monitoring
Function
The System monitoring enables to monitor system errors.
In order to set up the System monitoring function for a relay output, the
relay output must not be linked with a sensor (see section 4.3).
It can be used to monitor the following system errors.
Settings
Collective error
message
In the case of a Collective error message, the relay opens if one of the
following malfunctions occurs:
One of the sensors properly registered at the controller does not
supply a valid main measured value
One of the sensors properly registered at the controller does not
supply a valid secondary measured value
The monitoring combi output module has not received any new data
from the controller for two minutes.
In any case, the relay remains open for 10 seconds and only closes
when the malfunction is no longer present.
In the following cases, the relay does not open despite an invalid
measured value:
Settings
Selection
Explanations
Power failure
On
Off
The Power failure On function
monitors the supply voltage in
the IQ S
ENSOR
N
ET
on the
combi output module. If the
voltage falls below the critical
value, the relay switches.
Communication
On
Off
The Communication On function
monitors the functioning of the
controller and the
communication to the combi
output module.
Collective error
message
On
Off
The Collective error message
On function simultaneously
monitors the proper functioning
of all sensors and the
functioning of the monitoring
combi output module. (for
details see below this table)
Action
Open
For all functions of the System
monitoring, the relay action is
set to Open.