beautypg.com

5 setting the relay outputs, Setting the relay outputs -15, Section 4.5 – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual

Page 187: T (see section 4.5

background image

MIQ/CR3; DIQ/CR3

Settings

4 - 15

ba76032e01

01/2012

4.5

Setting the relay outputs

The following steps describe operation with the 2020 XT system. With
the System 182 XT, the outputs of the DIQ/CR3 can be found directly
under the Settings menu (see system operating manual of the DIQ/
S 182 XT).

Fig. 4-15 150 - Settings of outputs and links

1

Call up the measured value display with

m

.

2

Open the Settings menu with

s

.

3

Highlight the Settings of outputs and links menu item with

d

and confirm with

g

. The Settings of outputs and links display

appears.

4

Highlight the Feature column with

d

and confirm with

g

.

5

Highlight a line for a relay output (Rx) in the Feature column
with

d

and confirm with

g

.

The Settings of outputs and links display opens.

6

Highlight the Relay function menu item with

d

and confirm

with

g

.

7

Select one of the functions in the following list with

d

and

confirm with

g

.