3 installation, 1 scope of delivery, 2 requirements of the measurement location – YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual
Page 21: 3 installation guidelines for lightning protection, Installation -1, Scope of delivery -1, Requirements of the measurement location -1, 3installation
System 182
Installation
3 - 1
ba76022e01
12/2011
3
Installation
3.1
Scope of delivery
The following parts are included in the scope of delivery of the
DIQ/S 182:
DIQ/S 182 Universal Transmitter
Accessory kit with:
– Contact carrier with screws
– ISO cap nuts with screws and ring washers
– Cable glands with sealing gaskets
Operating manual.
3.2
Requirements of the measurement location
The measurement location must meet the environmental conditions
specified in section 9.1 IQ S
3.3
Installation guidelines for lightning protection
When using the IQ S
ENSOR
N
ET
instrumentation, particularly in outdoor
areas, adequate protection against (electrical) surges must be
provided. A surge is a summation phenomenon of surge voltage and
surge current. It is generated through the indirect effect of a lightning
event or switching operation in the mains, in the grounding system and
in information technology lines.
To be adequately protected against the damaging effects of surges, an
integrated concept of the following protective measures is required:
internal device-related protective measures and
external protective measures of the installation environment.
The internal device-related protective measures are already integrated
in the YSI online measuring technology as so-called 'lightning
protection' (see chapter 9 T
).
The external protective measures of the installation environment can
be carried out with respect to the following guidelines:
1 All lines of the IQ S
ENSOR
N
ET
system must be
a) installed inside (or else close to) the grounded metallic mounting
constructions, e.g. handrails, pipes and posts if possible