MIQ/CR3; DIQ/CR3

Settings

4 - 9

ba76032e01

01/2012

Characteristic curves

Through the selection of the Start value and End value, the proportional
output can be operated with a positive or negative characteristic curve.

Positive characteristic curve:
Select the End value to be greater than the Start value.
The turn-on duration or frequency increases with an increasing
measured value (see page 4-10).

Negative characteristic curve:
Select the End value to be smaller than the Start value.
The turn-on duration or frequency decreases with an increasing
measured value (see page 4-11).

The maximum values for the pulse width or switching frequency are
assigned to the End value and the minimum values for turn-on duration
or frequency are assigned to the Start value.

Fig. 4-6

Positive characteristic curve

Fig. 4-7

Negative characteristic curve

Measured value

Switching frequency f or
Pulse

v

width

End value

Start value

Max.

Min.

Measured value

Start value

End value

Max.

Min.

Switching frequency f or
Pulse

v

width