MIQ/CR3; DIQ/CR3

Settings

4 - 5

ba76032e01

01/2012

4.1.3

Proportional output

In the case of proportional output, a relay switches cyclically on and off
in a defined measured value range (proportional range). At the same
time, the relay switches with a:

duration of operation that corresponds to the measured value
(pulse-width output, see page 4-7) or

switching frequency (frequency output, see page 4-8).

Proportional outputs can be used in the following way:

Output with one relay:
An output range is defined with a Start value and an End value. No
output takes place above and below the output range
(see page 4-6).

Output with two relays:
An output range is defined for each relay with a Start value and an
End value. One relay outputs in the upper output range and a further
relay in the lower output range (see page 4-6).