MIQ/CR3; DIQ/CR3

Settings

4 - 21

ba76032e01

01/2012

4.5.5

Frequency controller

Function

The characteristic of the frequency output is laid down in the Start
value
, End value, Frequency (f) min. and Frequency (f) max. settings.
The fundamentals of the function are described in the introductory
chapter (see section 4.1.3).

In order to set up the Frequency controller function for a relay output,
the relay output must be linked with a sensor (see section 4.3).

Settings

Characteristic curve

If a value is entered for End value that is greater than the Start value,
the output has a positive characteristic curve.

In order to obtain a negative characteristic curve, a value must be
entered for End value that is smaller than the value for Start value.

Setting

Selection/Values

Explanation

Measured variable

Main variable
Adjoining variable

Main variable
designates the
actual measured
parameter of the
sensor (e.g. pH,
oxygen, etc.).
Adjoining variable
designates an
additional
measured
parameter (e.g.
temperature).

Start value

within the
measuring range
(sensor-
dependent)

Minimum spacing:
5 % of the
measuring range

End value

Frequency (f) min.

0 to 120 1/min

Minimum spacing:
10 1/min

Frequency (f) max.

Frequency with error

0 to 120 1/min

In case of system
errors or sensor
errors (see page 4-
39),
the relay
switches with the
frequency
specified.

Action

Open
Close

Relay action (see
section 4.5.1)