beautypg.com

YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual

Page 49

background image

System 182

Installation

3 - 29

ba76022e01

12/2011

General installation

instructions

Observe the following points when attaching connecting wires to the
terminal strip

 Shorten all wires to be used to the length required for the installation
 Always fit all the ends of the wires with wire end sleeves before

connecting them to the terminal strip

 Any wires that are not used and project into the enclosure must be

cut off as closely as possible to the cable gland.

 Screw a small cable gland with sealing ring into each remaining free

opening and close it with a blind plug.

Warning

No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. This could result in life
threatening electric shock when working with the IQ S

ENSOR

N

ET

.

Always cut off any wires that are not in use as closely as possible
to the cable gland.