3: Using the Curie System

29

If the signal you collect . . .

You can . . .

By . . .

Has too little light

Increase the light level on
scale in Scope mode

Increasing the integration time, or

Removing LVFs from the light path

Examples of Saturated Signal

Saturated Fluorescein Signal in Scope Mode

640-11100-000-02-0406