Configuring data acquisition parameters – Ocean Optics Curie UV-VIS Emission User Manual
Page 22
2: Configuration
• Method 2: Check the Strobe/Lamp Enable box on the Acquisition Parameters toolbar.
Once you enable the strobe via any of the options listed above, the lamp activates and a trace displays on
the graph when the CVD-DIFFUSE is placed in the cuvette holder. See
more detailed information on using the CVD-DIFFUSE with your Curie system. Adding a sample in the
cuvette holder should change the graph accordingly. If these results occur, your hardware and software
have been installed correctly.
Configuring Data Acquisition Parameters
►
Procedure
1. Using OOIBase32, select Spectrum | Configure Data Acquisition from the menu or click the
button. The Configure Data Acquisition screen appears with the Basic tab selected.
2. On
the
Basic tab, configure the integration time, averaging, and boxcar smoothing values. See the
OOIBase32 Spectrometer Operating Software, Operating Instructions for more information. See
Performing a Fluorescence Experiment
for more detailed information on choosing the optimal
data acquisition parameters for your measurement.