Spectrometer configuration dialog box – Ocean Optics Curie UV-VIS Emission User Manual

Page 17

background image

2: Configuration

3. In the USB Serial Number field, select the serial number of the USB2000 spectrometer in your

Curie system from the drop-down list.

4. Click OK.

Spectrometer Configuration Dialog Box

Procedure

To set the system parameters, do the following:

1. Select

Spectrometer | Configure from the menu. The Spectrometer Configuration screen

appears.

2. On the Wavelength Calibration tab, ensure that the calibration coefficients (read from a memory

chip in the Curie spectrometer) match the coefficients listed on the Wavelength Calibration data
sheet that came with your spectrometer.

3. Ensure that both the Master and Channel Enabled are selected.

4. Select the A/D Interface tab.