23-3 in case of ix81 – Olympus IX2-DSU BX-DSU User Manual
Page 80

ASSEMBLY
/ Z board setting and installation
II.
DSU
II .
8 - 2 6
Page
Z board installation
You must install Z board after you turn power of control box
to OFF.
1. Loosen fixing knobs (6 pieces) for two slot covers at back side of control
box (1) and remove covers and knobs.
2. Aligning direction of connector of control box and connector of Z board
(2), insert board along with board rails.
Push in until connectors are firmly connected.
3. Using knobs you removed, fix Z board (2). Likewise, attach the other
cover (3).
Keep one piece of slot cover that remained with care as it
may be required for Z board when it is removed.
8-23-3 In case of IX81
For Z board, use type U-ZPCB(T2) for sure.
In case that U-ZPCB(T2) or subsequent is not used, the following
troubles may occur even if correct setting was done.
● Rotation of focusing handle and actual objective lens movement would
be reversed.
● Focusing control may not be able to initialize, depending upon the
position. (Microscope does not boot up).
All DIP switches on Z board, S1, S2 and S3 are turned to OFF at ex-factory.
Note
Note
Note
TIP
(1)
(2)
(3)