2 installation of light illuminator, 2-1 in case of bx, Assembly – Olympus IX2-DSU BX-DSU User Manual
Page 57: 8 - 3 page
ASSEMBLY
/ Installation of light illuminator
II.
DSU
II .
8 - 3
Page
8-2 Installation of light illuminator
8-2-1 In case of BX
1.
Remove caps of mounting screws (2) on top of light illuminator (1)
with use of tweezers.
2. When using BX61, BX62, BX51 or BX52, put arm lift kit (WI-
ARMAD) on the arm and fix mounting screws (3) (4 pieces) to the
light illuminator firmly, using dedicated
the Allen wrench that
comes with kit.
3. When mounting the light illuminator, put it over slightly forward
position without aligning mounting screws (2) with mounting holes
(4) of microscope body.
4. Next, pressing the light illuminator towards right side and press it
towards rear side so that it would be positioned in correct
mounting position.
5.
While pressing the light illuminator towards rear side, fix mounting
screws (2) (4 pieces) to the light illuminator by turning the
dedicated
the Allen wrench that comes with kit in clockwise
direction.
6.
Attach caps of mounting screws (2) to their original positions.
7. When using BX61, BX62, BX51 or BX52, revolving nosepiece
adapter (DSU-READ) must be attached as follows.
1)
Loosen fixing screw (5) of microscope frame with use of the
dedicated the Allen screwdriver.
2) Align
revolving nosepiece adapter (6) with mount dovetail
from front side of microscope frame and push it and then,
tighten fixing screw (5).
3) Loosen fixing screw (7)
located at right side of revolving
nosepiece adapter, using the Allen screwdriver.
4) Align revolving nosepiece (8) with mount dovetail from front
side of microscope frame and push it and then, tighten fixing
screw (7).
(2)
(4)
(1)
(3)
(5)
(6)
(7)
(8)