ASSEMBLY

/ Lamp power supply setting

II.

DSU

II .

8 - 9

Page

8-8 Lamp power supply setting

Cord or cable is weak against bend or twist. Care must be

taken so as not to apply excessive force.

When connecting power cord, turn main siwtch to

mark

side (OFF).

Use power cord that comes our accessory for sure.

1. Verify that voltage and frequency of input power are within the range as

shown in identifying plate (1).

(The 100 V system can be used with voltages in the 100 to 120 V range

and 200 V system can be used with voltages in the 220 V to 240 V range,

both with frequencies of 50 or 60 Hz.)

2.

Connect connecting cord to connector (2) of power supply unit.

3. Install power cord to input receptacle (3) of power supply unit and insert

power plug (4) into power consent (5).

Power cord should be connected to 3-core consent with

earth terminal. When the earth of consent is not connected,

the electric safety performance intended by our company

may not be acquired.

(1)

(2)

(3)

(4)

(5)