beautypg.com

12 adjustment of field iris diaphragm, Details of each operation, 3 - 1 2 page – Olympus IX2-DSU BX-DSU User Manual

Page 36

background image

DETAILS OF EACH OPERATION

/ Adjustment of field iris diaphragm

II.

DSU

II .

3 - 1 2

Page

3-12 Adjustment of field iris diaphragm

1.

Turn front side turret (1) and insert fluorescent mirror unit desired

into light path.

2.

Turn rear side turret and insert reference mirror of No. 1 into light

path.

3.

Set objective lens of 10x and set the specimen.

4. Move shutter knob (2) located at lower right side of turret to left

(

mark side) to open the shutter.

5.

Use coarse/fine handle or, in case of motorized microscope, use

focus handle unit or focusing button on microscope frame to

bring the specimen in focus.

6. Pull out field iris diaphragm lever (3) located at side of

illumination tube and squeeze it to the smallest diameter.

7. Insert the Allen screwdriver that comes as accessory into

centering hole (4) of microscope frame located at side of

illumination tube (2 places) and adjust the axis of field iris

diaphragm image in a way that it comes in the same axis with the

filed of view.

8. Push in field iris diaphragm lever (3) and open field iris

diaphragm until it inscribes the field of view. In case that the axis

is out of center, do centering procedure again.

9. Push in field iris diaphragm lever (3) further and make the

aperture to circumscribe with field of view.

(In case of BX)

(In case of IX)

(1)

(1)

(2)

(3)

(4)

(2)