TROUBLESHOOTING GUIDE
II.
DSU
II .
5 - 1
Page
5 TROUBLESHOOTING GUIDE
Depending upon how the system is used, it may not exhibit adequate performance of the system, though it may
not be a failure. When trouble occurs, exercise adequate countermeasure with reference to the following table.
In case that trouble has not been improved, please contact local Olympus representative for assistance.
Trouble
Cause
Countermeasure
Ref.
Cable is not connected
correctly.
Connect cable correctly.
8-1
Disk unit is not correctly
inserted into disk box.
Insert disk unit into correct position firmly.
8-13
Interlock function is
activated as access panel
is not closed correctly.
Attach access panel firmly.
8-13
Disk does not turn.
• Stripes of disk appear on
the monitor.
• Confocal effect cannot
be acquired even if disk
were put into light path.
Transport lock is left as
attached.
Remove transport lock.
8-4
Mirror unit is about to
come off the cassette.
Remove turret from mirror unit cassette and
fix it to correct position firmly.
8-12
Camera mirror unit does
not turn.
Electric filter wheel does
not turn.
Cable is not connected
correctly.
Connect cable correctly.
8-1
Power of camera is not
turned ON.
Switch on power of camera.
3-5
Shutter is entering light
path.
Displace shutter from light path.
3-9
Filter like that is entering
light path.
Displace unnecessary substance from light
path.
3-11
Specimen image does not
appear on the monitor.
Combination of visual
mirror unit and camera
mirror unit is incorrect.
Set visual mirror unit (front side) to
reference mirror and set camera mirror unit
(rear side) to mirror unit.
3-10
3-7