VARIETIES OF OBSERVATION METHODS

/ Visual observation (BX)

II.

DSU

II .

4 - 3

Page

When change of reflected-light fluorescence and transmitting Nomarski

differential interference is frequently done and simultaneous observation

is also required, it is recommended that high abrasion-resistant analyzer

slider (U-ANH) is used instead of analyzer (U-AN).

When change of reflected-light fluorescence and transmitting Nomarski

differential interference is frequently done but simultaneous observation

is not required, it would be convenient if mirror unit (U-MDICT3) for

differential interference is attached at front side turret, instead of using

analyzer (U-AN or U-ANH).

If the fluorescent mirror unit is changed to differential interference mirror

unit, the analyzer comes into light path together so that the changeover is

easy.