ASSEMBLY

/ Z board setting and installation

II.

DSU

II .

8 - 2 5

Page

3. Using knobs you removed, fix Z board (2). Likewise, attach the other

cover (3).

Keep one piece of slot cover that remained with care as it

may be required for Z board when it is removed.

8-23-2 In case of BX61W1

For Z board, use type U-ZPCB(T2) for sure.

In case that U-ZPCB(T2) or subsequent is not used, the following

troubles may occur even if correct setting was done.
● Rotation of focusing handle and actual objective lens movement would

be reversed.

● Focusing control may not be able to initialize, depending upon the

position. (Microscope does not boot up).

All DIP switches on Z board, S1, S2 and S3 are turned to OFF at ex-factory.

Setting change of ONBOARD DIP SW

Use DIP SW set to OFF except the following DIP SW.

Turn DIP SW that is to be changed only to ON .

S2 Changes NO.2 and NO.3 to ON.

S3 Changes NO.2, NO.4 and NO.5 to ON.

If DIP switches are not correctly set, objective lens comes

down and hits the specimen.