STARTING
Prior to starting the device, the system and process-specific safety measures must be observed.
For the media to be supplied, the relevant safety requirements and measures must be taken into
account.
W A R N I N G !
Prior to starting the device, it must be verified that the system
voltage corresponds with the voltage shown on the rating plate!
N O T E !
If the set temperature on the capillary controller should be reduced
in one step by more than 30°C, the excess temperature circuit
activates the thermostat (press reset button to switch on again).
The following step-by-step procedure is recommended:
Check the set temperature on the built-in thermostat or on the external controller.
W A R N I N G !
Caution: Do not touch the probe surface when in use. The high
surface temperatures that prevail are liable to cause burns.
Protective gloves must be worn and the probe must be protected to
prevent unauthorised access!
Switch on the power.
N O T E !
The total heating time is about 40 minutes. The probe is ready to
use after exceeding the lower alarm threshold value (30°C below
the set value).
C A R E !
In the event of low temperature (failure of the probe heating), the
sample gas supply must be interrupted by suitable measures!