Gas sampling and gas conditioning technology

2-1.1.5-ME

12.5.1 VERSIONS WITH INTERNAL CAPILLARY TUBE THERMOSTAT

Remove the connection box cover. The electrical wiring
diagram shown is contained in the cover.

Insert the mains cable (min. 3 x 1.5 mm

2

) through the

cable gland and connect to the appropriate terminals.

Insert the signal cable (low temperature alarm) through
the cable entry and connect to the appropriate terminals
(contact position T

u

indicates alarm).

Screw the cover back in place again.








N O T E !


The use of temperature-resistant cable is recommended!


12.5.2 VERSION WITH EXTERNAL THERMOSTAT

Remove the cover of the connection box. The

electrical wiring diagram shown is contained in
the cover.

Insert the mains cable (min. 3 x 1.5 mm

2

)

through the cable gland and connect to the
appropriate terminals.

Insert the temperature sensor cable through

the cable entry and connect to the appropriate
terminals.

Screw the cover back in place again.





N O T E !


Use appropriate compensating cables for thermocouples!



Power 230V 50Hz
or 115V 60Hz

Alarm output

1

2

1

2

+30

-30

800W

Power 230V 50Hz
or 115V 60Hz

PT100 or
thermocouple

1

2

1

2

800W