Trouble shooting – M&C TechGroup MP-F10_R Operator's manual User Manual

Page 17

background image

10.1a-ME

Gas sampling and gas conditioning technology

17

15

TROUBLE SHOOTING

Before working on the pump isolate the power supply securely, then check that the lines are not live.

Problem/Indication

Possible cause

Check/Action

Pump produces no
flow

No main supply

Check power supply; Check plug for correct
fit

Connections or lines are
blocked

Remove blockade

An external valve is closed or
a filter is blocked

Open valve or clean dirty/blocked filter

Liquid (condensate) has col-
lected in the pump head

Let pump convey air a few minutes

Flow, pressure or
vakuum too low

Diaphragm or valves are worn Change diaphragm or/and valves

Compare the actual perform-
ance with the figures in chap-
ter 9

The pump is not designed for this condition

There is pressure on the
pressure side and at the same
time vacuum on the suction
side

The pump is not designed for this condition

The

cross-section

of

the

pneumatic lines or connected
components is too small, or
they are restricted

To measure the performance, disconnect the
pump from the system (smaller diameter
tubing or a valve can significantly affect
performance)

A leak at the connectors, lines
or pump head. Bellows or
valve plate is damaged or
pump head is contaminated

Insulate the leak, tighten the screws, clean
or exchange contaminated parts