Trouble shooting – M&C TechGroup MP-F10_R Operator's manual User Manual
Page 17

10.1a-ME
Gas sampling and gas conditioning technology
17
15
TROUBLE SHOOTING
Before working on the pump isolate the power supply securely, then check that the lines are not live.
Problem/Indication
Possible cause
Check/Action
Pump produces no
flow
No main supply
Check power supply; Check plug for correct
fit
Connections or lines are
blocked
Remove blockade
An external valve is closed or
a filter is blocked
Open valve or clean dirty/blocked filter
Liquid (condensate) has col-
lected in the pump head
Let pump convey air a few minutes
Flow, pressure or
vakuum too low
Diaphragm or valves are worn Change diaphragm or/and valves
Compare the actual perform-
ance with the figures in chap-
ter 9
The pump is not designed for this condition
There is pressure on the
pressure side and at the same
time vacuum on the suction
side
The pump is not designed for this condition
The
cross-section
of
the
pneumatic lines or connected
components is too small, or
they are restricted
To measure the performance, disconnect the
pump from the system (smaller diameter
tubing or a valve can significantly affect
performance)
A leak at the connectors, lines
or pump head. Bellows or
valve plate is damaged or
pump head is contaminated
Insulate the leak, tighten the screws, clean
or exchange contaminated parts