beautypg.com

Receipt of goods and storage, Installation instructions – M&C TechGroup MP-F10_R Operator's manual User Manual

Page 10

background image

10

Gas sampling and gas conditioning technology

10.1a-ME

10

RECEIPT OF GOODS AND STORAGE

Please take the bellows pump and possible special accessories carefully out of the packaging
material immediately after arrival, and compare the goods with the items listed on the delivery
note!

Check the goods for any damage caused during delivery and, if necessary, notify your trans-
port insurance company without delay of any damage discovered.

N O T E !


The bellows pump should be stored in a protected, frost-free room!



11

INSTALLATION INSTRUCTIONS

When installing the pump make certain that accident prevention regulations and safety instructions
including those for subsequent operation are observed. The safety instructions in section 6.2 must be
observed.

N O T E !

The pump must only be used under conditions specified in the
technical data. The pump should be installed away from heat
sources and freely ventilated to prevent any accumulation of heat.
For outdoor installation, the pump must be installed in a housing
protected from frost in the winter and sufficiently ventilated in
summer. Exposure to direct sunlight must be avoided.


The protection class of the pumps MP-F... is IP55 and so offer no protection against contact or foreign
bodies.

W A R N I N G !



It is therefore essential to provide protection for persons against
contact with live parts (e.g. electrical connections, motor windings)
and moving parts (e.g. fan). Protection against entry of foreign
bodies must also be provided.

W A R N I N G !

Pumps have mechanical moving parts that can induce vibrations.
To prevent damages at the pump or at peripheral components /
facilities as well as minimizing noise development an appropriate
vibration decoupling is necessary. For this M&C can deliver e.g.
anti-vibration pads.
This is also explicit valid for the connection of the sample lines at
the pump head.

W A R N I N G !

The components connected to the pumps have to be designed for
the pneumatic data of the pumps.
Observe the corresponding safety regulations when connecting the
pumps to the mains.
For the media to be pumped the corresponding safety regulations
have to be observed.