Introduction, Application, Figure 1 – M&C TechGroup ECS Operator's manual User Manual
Page 8: Example of application of ec/ecs

8
Gas sampling and gas conditioning technology
3-4.1-ME
6
INTRODUCTION
The patented M&C EC/ECS gas cooler unit is always to be installed in situations where there is
interference from moisture in the gas to be measured.
Reduction of the gas temperature inside the cooler to a stable and very low dew point effects a
condensing-out of the sample gas.
7
APPLICATION
Figure 1 shows a typical example of an application for installation of an EC/ECS gas cooler unit.
Sample gas
in
+5°C
EC
Collecting vessel
Check-
gas in
Filter sample probe SP...
3-way ball valve
Super-fine filter FP...
Heated sample line
Peristaltic pump
Aerosol filter CLF
EC/ECS cooler
Membrane pump
Flow meter FM 10
Analyser
Figure 1 Example of application of EC/ECS
The gas to be measured is extracted with a gas sample pump via a gas sample probe and a
heated sample line and cooled down in the gas cooler EC/ECS to a dew point of +5°C. The
emerging condensate is removed by a peristaltic pump or a collecting vessel . The super-fine filter
located afterwards removes solid particles. For increased operating safety of the entire system we
recommend installing a super-fine filter with a liquid alarm sensor. If required an aerosol filter
can be installed in front of the flow meter . The gas thus treated can now be passed into the
analyser . For calibration of the analyser a ball-valve is existing.