Electrical connections – M&C TechGroup ECS Operator's manual User Manual
Page 32
32
Gas sampling and gas conditioning technology
3-4.1-ME
For mounting the display and the temperature sensor please notice the following:
Loosen up the fixing screws of the front panel and remove it (see operating manual EC/ECS);
Remove the cooler cover (see operating manual EC/ECS);
Make the cut-out for the temperature display into the front panel according to the locating distance
shown in figure 13;
For cable bushings through the plastic housing of the EC/ECS control board please use the
available PG9 cable glands.
21.3
ELECTRICAL CONNECTIONS
N O T E !
Attention must be paid to the requirements of VDE 0100 when setting UP
high-power electrical units with nominal voltages of up to 1000V together
with the associated standards and stipulations.
Check the details on the type plate to ensure that the cooler is connected up
to the correct mains voltage!
The electrical connection of the digital temperature display happens on the EC/ECS control board (see
operation manual EC/ECS).
The temperature display is already equipped with all necessary cables.
The following steps must be done:
Loosen up the fixing screws of the cover of the plastic housing and remove the cover;
The cables of the digital temperature display have to be connected as follows:
digital temperature display
cable [colour]
Meaning
8-channel connector
clamp point
yellow
+ 10V
11
brown
0V
12
green
1mV/°C
9
white
0V
12