Gas sampling and gas conditioning technology

3-4.1-ME

Condensate removal is done by customer according to the type of operation with:

- Automatic float-type condensate traps AD-... only for over-pressure operation;

N O T E !

Stainless steel heat exchangers with G 3/8“ thread joint can be directly fitted
up with the float-type condensate trap AD-SS by means of a thread adapter
part number FF 11000

(1/2“ NPT to G 3/8“i). By this wall mounting of the AD-

SS unit isn’t necessary!

- Condensate collector container that is emptied manually;

- peristaltic pump.

Option: automatic condensate removal with peristaltic pump SR25.1 resp. SR25.2

In case of peristaltic pumps type SR25.2, build in the front plate, the condensate outlet 4/6 mm hose
is provided directly at the corresponding pump.

The Universal unit EC-FD with peristaltic pumps SR25.1 is equipped with a condensate outlet for
4/6mm hose connection on the front panel (see fig. 2).

13.2

ELECTRICAL CONNECTIONS

W A R N I N G !

When connecting the equipment, please ensure that the supply
voltage is identical with the information provided on the model
type plate.

N O T E !

Attention must be paid to the requirements of IEC 364 (DIN VDE
0100) when setting high-power electrical units with nominal
voltages of up to 1000 V, together with the associated standards
and stipulations.

Check the details on the type plate to ensure that the equipment is
connected up to the correct mains voltage.

The main circuit must be equipped with a fuse of 10A

T

(over

current protection); for electrical details see technical data.