Remote Interface Reference
Error Messages
•
-103, "Non-boolean argument"
The command required a parameter in the form of 0 or 1. No other form of
the parameter is allowed.
•
-104, "Missing parameter"
The command required at least one argument which was not found before
the termination character(s).
•
-105, "Out of range"
At least one of the parameter values received was out of the valid range.
Refer to the Menu Reference section beginning on page 43 for the
acceptable ranges for each instrument setting.
•
-106, "Undefined length"
The user attempted to set a parameter before a length value was specified
for the assigned calibration.
•
-107, "Percent units for length"
The user attempted to set the active length in the remote units of percent.
The remote units must be inches or centimeters to set the active length.
•
-108, "Incorrect passcode"
The user attempted to unlock the calibration and provided the incorrect
passcode for the
CAL:UNLOCK
command.
•
-109, "Command not supported"
The user sent a command to the unit that is not supported.
•
-110, "Cal is locked"
The user attempted to modify a parameter associated with a calibration of
a sensor input but failed to first unlock the calibration. See the
CAL:UNLOCK
command.
•
-111, "Invalid offset factor"
The user sent an offset factor that was out of the acceptable range of
−
100% to +100%, or would result in an overflow of the internal
measurement circuits.
•
-112, "Invalid approx factor"
The user sent an approximate calibration factor that was out of the
acceptable range of 10% to 1000%, or would result in an overflow of the
internal measurement circuits.