93

Service

Line Voltage Selection

receptacle. Failure to do this could expose the user to high voltages
and could result in life-threatening electrical shock.

Caution

The Model 286 instrument operates on 50-60 Hz power and may be
configured for 100-120 or 200-240 VAC ±10%. The power
requirements for each instrument are marked on the rear panel. Be
sure the instrument’s power requirements match your power source
prior to plugging in the line cord. Do not fail to connect the input
ground terminal securely to an external earth ground.

If the instrument operating voltage range needs to be changed, ensure the
instrument is de-energized by disconnecting the power cord from the
power source. Remove the instrument cover and slide the voltage selector
switch (S1) on the main printed circuit board to the proper voltage; for 100
to 120 VAC input “115V” should be visible in the switch window and for
200 to 240 VAC input “230V” should be visible in the switch window.
Replace the instrument cover and indelibly mark the rear panel
indications to match the new configuration
.

6.4 Additional Technical Support

If the cause of a problem cannot be located, contact an AMI Technical
Support Representative at (865) 482-1056 for assistance. The AMI
technical support group may also be reached by Internet e-mail at

[email protected]

. Additional technical information,

latest software releases, etc. are available at the AMI World Wide Web site
at:

http://www.americanmagnetics.com

Do not return the Model 286 or other AMI system components to AMI
without prior return authorization.

6.5 Return Authorization

Items to be returned to AMI for repair (warranty or otherwise) require a
return authorization number to ensure your order will receive proper
attention. Please call an AMI representative at (865) 482-1056 for a return
authorization number before shipping any item back to the factory.