Calibration – American Magnetics 286 Multi-Sensor Liquid Level Instrument (CE-Marked) User Manual

Page 34

background image

20

Calibration

Calibration Methods

Figure 3-2. Calibration method selection diagram.

Instrument &

sensor purchased

together?

START

Verify sensor

specifications

Verify calibration

sticker on bottom of

instrument

Perform

Open

Dewar Calibration

Is factory

calibration

correct?

Can

Open

Dewar Calibration be

performed in target

liquid?

Can

Closed

Dewar Calibration be

performed in target

liquid?

Perform

Closed

Dewar Calibration

Perform

Approximate

Calibration

Go to

Operation

Y

N

N

Y

N

Is factory

calibration

Approximate?

Y

Y**

N

** Factory Approximate Calibration should only be used if
open dewar or closed dewar calibration is not feasible in
the target liquid at the customer's facility.

N

Can

Open

Dewar Calibration be

performed in target

liquid?

Can

Closed

Dewar Calibration be

performed in target

liquid?

Y

Y

N

N

Y