Calibration
Closed dewar calibration
9. Calculate the factor C
adj
using the following equation:
where L
total
is the total sensor length in inches, L
active
is the active
sensor length in inches, and
ε
is the dielectric constant of the target
liquid.
10. Enter C
adj
into the instrument by placing the instrument in the
custom approximate calibration factor entry screen as documented
in paragraph 3.3.4.2 on page 31.
It will be necessary to
ESC
to the sensor input selection screen and
choose the
Select
option in order to traverse the calibration menu
tree and arrive at the custom approximate calibration factor entry
screen. Ensure that the same calibration used to this point of the
presetting procedure is selected for the entry of a custom
approximate calibration factor.
The presetting procedure is complete. The presetting procedure has
temporarily calibrated a measurement span that should prevent the fill for
a MAX calibration from exceeding a 100% indication from the Model 286.
Proceed to the remainder of the closed dewar calibration procedure as
presented below.
3.2.2.2 Completing the closed dewar calibration procedure
1. Install the sensor in the dewar and energize the instrument with
the sensor appropriately connected to the instrument (see the
system diagram on page 5).
2. Again, re-enter the calibration menu by using the passcode if
necessary, and select the identical sensor input to calibrate as
selected in the presetting procedure.
3. Choose the
Perform
option and select the identical calibration as
selected for the presetting procedure.
4. Verify the active length value previously entered in the presetting
procedure.
5. Place the instrument in the MIN/MAX entry screen as documented
in paragraph 3.3.5.3 on page 34.
6. Connect a strip chart recorder to the 0-10 V recorder output
terminals available on the rear panel of the instrument if an analog
output option was purchased. If the recorder output is not
C
adj
120 1
ε
1
–
(
)
L
active
L
total
---------------
+
=