beautypg.com

3 calibration – American Magnetics 286 Multi-Sensor Liquid Level Instrument (CE-Marked) User Manual

Page 31

background image

17

3 Calibration

The Model 286 Liquid Level Controller is typically calibrated at the factory
for specific length sensors for use in specific target liquids. The calibration
lengths and calibration liquids are listed on the configuration sheet
supplied with this manual. If a factory calibration method utilized was
approximate, the calibration length will be noted as an approximate value.

3.1 Calibration Concepts

3.1.1

Relationship between Calibration and Sensor Length

The capacitance-based method of measuring the liquid level operates by
measuring the period of an oscillator, which is contained in the oscillator/
transmitter unit (or contained in the instrument for sensor inputs A and
B). As the liquid level varies, the value of the capacitance and the period of
the oscillator vary proportionally. Since the dielectric properties of liquids
vary and the component tolerances for the sensor and oscillator introduce
variations, a calibration is required to assure maximum accuracy for a
specific sensor immersed in the target liquid. The calibration MIN and
MAX settings correspond to the minimum and maximum oscillation
periods, respectively, for a given sensor and target liquid configuration.

The LENGTH setting of the instrument is only provided as a means of
scaling the 0% (MIN) to 100% (MAX) range of the measurement to
meaningful units of length. Before the calibration it is important to
accurately measure the distance between the physical locations on the
sensor corresponding to the desired MAX and MIN calibration points. The
measured value for the length will be used in configuring the instrument
for operation.

3.1.2

Variations in the Dielectric with Changing Density

For cryogenic liquids, the dielectric of the liquid will change with a change
in density. The amount of change is dependent on the properties of the
specific liquid. Figure 3-1 illustrates the variations in dielectric for
nitrogen vs. pressure under saturated conditions.

1

Since the instrument

uses a capacitance-based method for determining liquid level, such a
change in the dielectric of the liquid will result in a shift in the level
reading of the instrument. The calibration procedures described herein are
most accurate when applied in situations where the operating conditions
of the cryo-vessel are relatively constant, i.e. the operating pressure and
temperature of the cryo-vessel are relatively constant.

1. Data obtained from NIST Standard Reference Database 12.