American Magnetics 286 Multi-Sensor Liquid Level Instrument (CE-Marked) User Manual
Page 94
![background image](/manuals/252950/94/background.png)
80
Remote Interface Reference
Remote Units Commands and Queries
•
{A|B|C|D}:CAL:APPROX <
selection
>,<
approx_factor
>
Sets the approximate calibration factor for the sensor input and
calibration selection specified.
•
{A|B|C|D}:CAL:APPROX <
selection
>?
Returns the approximate calibration factor for the sensor input and
calibration selection specified.
•
{A|B|C|D}:CAL:OFFSET <
selection
>,<
offset_factor
>
Sets the offset factor for the sensor input and calibration selection
specified.
•
{A|B|C|D}:CAL:OFFSET <
selection
>?
Returns the offset factor for the sensor input and calibration selection
specified.
•
CAL:LOCK
Prevents any subsequent commands from modifying any sensor input
calibration. Queries related to the calibration may still be performed.
5.4.8
Remote Units Commands and Queries
The remote units commands and queries provide the capability of selecting
between units of percent, inches, and centimeters for all parameters and
return values for level and alarm commands and queries. The remote units
are independent of the display units selected via the front panel. The
remote units setting is persistent between power-ups of the Model 286.
•
UNITs {0|1|2}
Selects between the percent (0), inches (1), and centimeters (2) units. For
example, the command:
UNITS 1;
selects inches for the remote units.
•
UNITs?
Returns the currently selected units.
0, ”PERCENT”
is returned if the remote units are percent.
1, ”INCHES”
is returned if the remote units are inches.
2, ”CM”
is returned if the remote units are centimeters.
•
PERCent
Sets the remote units to percent.