beautypg.com

Power – American Magnetics 286 Multi-Sensor Liquid Level Instrument (CE-Marked) User Manual

Page 102

background image

88

Service

Troubleshooting Hints

9. Use only grounded-tip soldering irons.

6.2.1 The Model 286 does not appear to be energized with the

power switch in the

POWER

(

I

) position.

1. Ensure that the Model 286 is energized from a power source of

proper voltage.

Warning

If the instrument has been found to have been connected to an
incorrect power source, return the instrument to AMI for evaluation
to determine the extent of the damage. Frequently, damage of this
kind is not visible and must be determined using test equipment.
Connecting the instrument to an incorrect power source could
damage the internal insulation and/or the ground conductors,
thereby, possibly presenting a severe life-threatening electrical
hazard.

2. Verify continuity of the line fuse, F1, located on the instrument

printed circuit board.

Warning

This procedure is to be performed only when the instrument is
completely de-energized by removing the power-cord from the power
receptacle. Failure to do so could result in personnel coming in
contact with high voltages capable of producing life-threatening
electrical shock.

a. Ensure the instrument is de-energized by disconnecting the

power cord from the power source. Disconnect the power cord
from the connector located on the rear panel of the instrument.

b. Remove the instrument top cover and check the fuse F1 for

continuity.

c. If the fuse is open, replace with a fuse of identical rating.

Caution

Installing fuses of incorrect values and ratings could result in
damage to the instrument in the event of component failure.

Note

The proper fuse for this instrument is a 315 mA IEC 127-2 Type F
sheet II 5x20mm. The fuse will be labeled with the current rating
and marked with an F to denote fast-acting.