Calibrating analog channels, Calibrating rtd – Perle Systems Perle IOLAN DS1 User Manual

Page 177

background image

177

Calibrating I/O

Calibrating RTD

When calibrating the IOLAN Analog input for RTD, you will need a resistor that is better than .05%
Ohms accuracy. When you calibrate one channel, all RTD channels are automatically calibrated for
that range; if another channel is set for a different range, you will need to calibrate that channel
separately, but all channels that use that range are automatically calibrated.

Calibrating Analog Channels

Analog Input can be calibrated for Analog and Temperature IOLAN models.

Select the channel you want to calibrate. This example uses an A4 model that has channel A1 set to
Current with a Range of 0 to 20mA.

If you have not disabled confirmation messages (

Tools

,

Options

in DeviceManager only), you will

get prompted to verify channel calibration.

Click

Yes

to proceed with calibration. You are now prompted to apply 0 mA to the positive (+) and

negative (-) terminals. Once that is done, click

Yes

to proceed.

You are now prompted to apply 20 mA to the positive (+) and negative (-) terminals. Once that is
done, click

Yes

to proceed.

Once calibration is successfully completed, click

OK

to finish the process.